Axis-Shield Homocysteine EIA

 

Axis-Shield Homocysteine EIA microtitre plate ELISA (96 Well)

Product Code: FHCY100

 

 

 

Assay Principle
Axis-Shield Homocysteine Enzyme Immunoassay (EIA) is an enzyme immunoassay for the determination of homocysteine (Hcy) in serum or plasma.

Protein-bound Hcy is reduced to free Hcy and enzymatically converted to S-adenosyl-L-homocysteine (SAH) in a separate procedure prior to the immunoassay. The following solid-phase ELISA is based on competition between SAH in the sample and immobilised SAH bound to the walls of the microtitre plate for binding sites on a monoclonal anti-SAH antibody. After a wash stage to remove unbound antibody, a secondary HRP-labelled rabbit anti-mouse antibody conjugate is added. The peroxidase activity is measured after addition of substrate and the absorbance is inversely related to the concentration of HCY in the sample.

 

 

Measuring Range
The calibrator range is from 2 to 50 µmol/L.

 

 

Assay Precision
Precision of the Axis-Shield Homocysteine EIA assay was evaluated according to NCCLS Protocol EP5. Three levels of controls were assayed for 20 days with 4 replicates per run at each level. Precision data are summarised in Table 1.

 

Table 1: Precision

 

Samples

Average HCY

µmol/L

 

Within-run

Precision

 

Total Precision

Control Low 6.1   8%   10%
Control Medium 10.5   7%   9%
Control High 20.6   8%   10%

 

 

 

Limit of Quantification
The quantification limit (CV < 20%) is 1.0 µmol/L.

 

 

Linearity of diluted plasma samples
If the homocysteine concentration of a sample exceeds the range of the calibration curve, the sample should be diluted with Reagent A and reanalysed.

The linearity was evaluated by diluting four high patient samples with varying amounts of Reagent A as diluent.

Linear regression analysis gave:
Slope: 0.98
Intercept: -0.4µmol/L
Correlation coefficient: r2: 0.99

 

 

Method Comparison
The Axis-Shield Homocysteine EIA was compared to the University of Bergen HPLC method. A comparison of 164 patient samples ranging from 3-37µmol/L homocysteine gave the linear regression shown below.
Slope: 0.94
Intercept: -0.09 µmol/L
Correlation coefficient: (r2): 0.94

 

 

Axis-Shield Homocysteine EIA (96 well) - FHCY100

Axis-Shield Homocysteine Control kit (tri-level, 1.5mL each) - FHCY200

 

Please visit www.axis-shield.com for information.

Homocysteine Testing

To find out more about the different Homocysteine assays available from Axis-Shield, visit www.axis-shield.com

NEW Instrument Applications

ABX PENTRA 400

MINDRAY BS-400 (SPIN 640)

 

Latest Publications

Homocysteine and Reclassification of Cardiovascular Disease Risk


 Associations Between Elevated Homocysteine, Cognitive Impairment, and Reduced White Matter Volume in Healthy Old Adults.


Towards the Identification of Blood Biomarkers for acute stroke in humans: a comprehensive systematic review.


For full details visit the Publications page.

© Axis-Shield plc. All rights reserved. Terms of Use | Accessibility
E: homocysteine@axis-shield.com - T: +44 13 8242 2000 - F: +44 1382 422088